Test voltage / test current |
Measuring range |
Resolution |
Accuracy |
50V 1-mA @ 50 kΩ |
0.01 ... 5 MΩ |
0.0001 MΩ |
± (4 % + 20 digits) |
|
5 ... 50 MΩ |
0.001 MΩ |
± (4 % + 20 digits) |
|
50 ... 500 MΩ |
0.01 MΩ |
± (4 % + 20 digits) |
|
500 ... 1000 MΩ |
0.1 MΩ |
not specified |
|
|
|
|
125V / 1-mA @ 125 kΩ |
0.01 ... 5 MΩ |
0.001 MΩ |
± (2 % + 20 digits) |
|
5 .... 50 MΩ |
0.001 MΩ |
± (2 % + 20 digits) |
|
50 ... 500 MΩ |
0.01 MΩ |
± (3 % + 20 digits) |
|
500 .... 5000 MΩ |
0.1 MΩ |
± (4 % + 20 digits) |
|
|
|
|
250V / 1-mA @ 250 kΩ |
0.01 ... 5 MΩ |
0.0001 MΩ |
± (2 % + 20 digits) |
|
5 ... 50 MΩ |
0.001 MΩ |
± (2 % + 20 digits) |
|
50 ... 500 MΩ |
0.01 MΩ |
± (3 % + 20 digits) |
|
500 ... 5000 MΩ |
0.1 MΩ |
± (4 % + 20 digits) |
|
|
|
|
500V / 1-mA @ 500 kΩ |
0.01 ... 5 MΩ |
0.0001 MΩ |
± (2 % + 20 digits) |
|
5 ... 50 MΩ |
0.001 MΩ |
± (2 % + 20 digits) |
|
50 ... 500 MΩ |
0.01 MΩ |
± (3 % + 20 digits) |
|
500 ... 5000 MΩ |
0.1 MΩ |
± (4 % + 20 digits) |
|
|
|
|
1000V / 1-mA @ 1000 kΩ |
0.01 ... 5 MΩ |
0.0001 MΩ |
± (2 % + 20 digits) |
|
5 ... 50 MΩ |
0.001 MΩ |
± (2 % + 20 digits) |
|
50 ... 500 MΩ |
0.01 MΩ |
± (3 % + 20 digits) |
|
500 ... 5000 MΩ |
0.1 MΩ |
± (4 % + 20 digits) |
|
|
|
|
Short circuit current |
<1.5-mA |
|
|
|
|
|
|
Multimeter |
Measuring range |
Resolution |
Accuracy |
Measuring parameters |
50 mV |
0.001 mV |
± (0.06 %+ 20 digits) |
DC |
500 mV |
0.01 mV |
± (0.06 %+ 6 digits) |
|
5 V |
0.0001 V |
± (0.06 %+ 4 digits) |
|
50 V |
0.001 V |
± (0.06 %+ 4 digits) |
|
500 V |
0.01 V |
± (0.06 %+ 4 digits) |
|
1000 V |
0.1 V |
± (0.1 %+ 3 digits) |
|
|
|
|
AC |
50 mV |
0.001 mV |
± (1 % + 50 digits) |
AC + DC |
500 mV |
0.01 mV |
± (1 % + 50 digits) |
50 .... 1000 Hz |
5 V |
0.0001 V |
± (1 % + 30 digits) |
|
50 V |
0.001 V |
± (1 % + 30 digits) |
|
500 V |
0.01 V |
± (1 % + 30 digits) |
|
1000 V |
0.1 V |
± (1 % + 30 digits) |
|
|
|
|
The specified accuracy for alternating voltage refers to 5 ... 100% of the measuring range
|
|
|
|
|
Direct current |
Measuring range |
Resolution |
Accuracy |
|
500 µA |
0.01 µA |
± (0.5 % + 3 digits) |
|
5000 µA |
0.1 µA |
± (0.5 % + 3 digits) |
|
50 mA |
0.001 mA |
± (0.5 % + 3 digits) |
|
500 mA |
0.01 A |
± (0.5 % + 3 digits) |
|
10 A |
0.001 A |
± (0.5 % + 3 digits) |
|
|
|
|
Alternating current |
500 µA |
0.01 µA |
± (1% + 30 digits) |
AC + DC |
5000 µA |
0.1 µA |
± (1% + 30 digits) |
50 .... 1000 Hz |
50 mA |
0.001 mA |
± (1% + 30 digits) |
|
500 mA |
0.01 A |
± (1% + 30 digits) |
|
10 A |
0.001 A |
± (1% + 30 digits) |
|
|
|
|
The specified accuracy for alternating current refers to 5 ... 100% of the measuring range
|
|
|
|
|
Resistance |
Measuring range |
Resolution |
Accuracy |
|
50 Ω |
0.01 Ω |
± (0.2 % + 20 digits) |
|
500 Ω |
0.1 Ω |
± (0.1 % + 10 digits) |
|
5 kΩ |
0.0001 kΩ |
± (0.1 % + 3 digits) |
|
50 kΩ |
0.001 kΩ |
± (0.1 % + 3 digits) |
|
500 kΩ |
0.01 kΩ |
± (0.1 % + 3 digits) |
|
5 MΩ |
0.001 MΩ |
± (1 % + 10 digits) |
|
50M Ω |
0.001 MΩ |
± (1.5 % + 50 digits) |
|
|
|
|
Capacitance |
50 nF |
0.001 nF |
± (1.5 % + 50 digits) |
|
50 nF |
0.01nF |
± (1.5 % + 10 digits) |
|
500 nF |
0.1 nF |
± (1.5 % + 10 digits) |
|
5 µF |
0.001µF |
± (1.5 % + 5 digits) |
|
50 µF |
0.01µF |
± (1.5 % + 5 digits) |
|
500 µF |
0.1µF |
± (1.5 % + 5 digits) |
|
5 mF |
0.001 mF |
± (3 % + 30 digits) |
|
10 mF |
0.01 mF |
± (3 % + 30 digits) |
|
|
|
|
Frequency |
50 Hz |
0.001 Hz |
± (0.02 % + 3 digits) |
(Electronics) |
500 Hz |
0.01 Hz |
± (0.02 % + 3 digits) |
|
5 kHz |
0.0001 kHz |
± (0.02 % + 3 digits) |
|
50 kHz |
0.001 kHz |
± (0.02 % + 3 digits) |
|
500 kHz |
0.01 kHz |
± (0.02 % + 3 digits) |
|
5 MHz |
0.0001 MHz |
± (0.02 % + 3 digits) |
|
50 MHz |
0.001 MHz |
± (0.02 % + 3 digits) |
|
100 MHz |
0.01 MHz |
not specified |
|
|
|
|
Sensitivity: 0.8 VRMS min @ 20% ... 80% duty cycle <100 kHz 5 V RMS min @ 20% ... 80%> 100 kHz
|
|
|
|
|
|
Measuring range |
Resolution |
Accuracy |
Frequency (Electronics) |
40 Hz ... 10 kHz |
0.01 1 Hz ... 0.001 kHz |
± (0.1%) |
Duty Cycle |
0.1 ... 99.9% |
0.01% |
± (1.2% + 2 digits) |
Pulse width |
100 µS ... 100 ms; Frequency: 5 Hz ... 150 kHz |
Temperature - |
58 ... 2462°F |
0.1°F |
± (0.5 % + 7°F) |
Type K Thermoelement |
-50 ... 13350 °C |
0.1°C |
± (0.5 % + 3.5°C) |
4 ... 20 mA% |
-25 ... 125% |
0.01% |
± 50 digits |
Resistance 50 Ω range |
50 Ω |
0.001 Ω |
± (1 % + 20 digits) |
|
500 Ω |
0.01 Ω |
± (1 % + 20 digits) |
|
|
|
|
Open circuit voltage: 5V; Overload protection 250V |
|
|
|
|
General information |
|
|
Memory |
9999 memory points |
Housing |
Double moulded housing, IP67 (waterproof and dustproof) |
Drop test |
2 m / 6ft, 6in |
Diode test |
Test current 0.9-mA |
|
Open circuit voltage 2.8-mA |
Continuity test |
Acoustic signal at resistance < 35 Ω / Test current < 0.35-mA |
Peak detection |
> 1ms |
Temperature sensor |
Thermoelement Type K |
Input impedance |
DC: > 10 MΩ |
|
AC: > 9 MΩ |
AC coupling |
TRMS |
Bandwidth AC voltage |
50 Hz ... 1 kHz |
Crest factor |
<3 at 500V |
|
< 1.5 at 1000V |
Display |
50000 digit LCD illuminated with bargraph |
Automatic shutdown |
after 15 minutes |
Measurement rate / update display |
10 Hz |
Power supply |
6 x 1.5V AA batteries |
Interface |
Wireless USB |
Fuses |
mA,µA: 0.5 A / 1000V fast ceramic |
|
A: 10 A / 1000V fast ceramic |
Operating conditions |
5 ... 40°C / 41 ... 104°F, max 80% r.H. |
Storage conditions |
-20 ... 60°C / -4 ... 140°F, max. 80% r.H. |
Operating altitude |
Max. 2000 m / 6561 feet |
Weight |
ca. 720 g / 1.6 lbs |
Dimensions |
220 x 95 x 50 mm / 8.6 x 3.7 x 1.9 in |
Safety standards |
EN61010-1 |
|
IEC 61010-1 Part 2 (2001) |
|
CAT IV 600V, CAT III 1000V |
|
UL 61010-1 Part 2 (2004) |
|
CAN/CSA C22.2 No 6110-1 Part 2 |
|
(2004) |
|
UL 61010B-2-031 Part 1 (2003) |